Contact Us Home Home

Special UV-Index Sensor ERYCA

Features

 The UVI Sensor ERYCA is a further development of our product EryF. With an additional filter the accordance with the erythema action curve of the human skin has been further improved. The ERYCA is designed for use as an erythema sensor according to ISO 17166 CIE S 007/E (2000) - DIN 5050. It is optimally suited to feature in high quality instruments for exact measurements of the UV Index.


EryF*

Special UVI type sensor, precision up to ± 0.5 UVI

Perfectly suited for high precision, low cost sun UV dosimetry

The Sensor's current is directly proportional to the UV-Index

Also suited for sun tanning bank dosimetry

Based on approved EryF technology

Schottky-type TiO2 photodiode

Intrinsic visible blindness due to wide-bandgap semiconductor material

Large photoactive area

Designed to operate in photovoltaic mode

· Hermetically sealed metal TO18 housing and UV-glass window

Maximum Ratings

Operating temperature range: -20 - +80 °C

Reverse voltage: 3V

Forward current: 1mA

Total power dissiption at 25°C: 1mW

General Characteristics

(Ta = 25 °C)

Active area: 4.18 mm²

Active viewing dimensions: 2.2 x 1.9 mm

Maximum viewing angle: 70 degrees

Shunt resistance (dark): 300 MOhms

Dark current at 10mV reverse bias: 30pA

Open circuit voltage (200mW/cm², λ=300nm): >250 mV

Min. Short circuit current (200mW/cm², λ=300nm): >160 nA

Breakdown voltage (dark): >3V

Spectral Characteristics

(Ta = 25 °C)

Min. spectral sensitivity: 19 mA/W

Wavelength of peak spectral sensitivity: 300 nm

Range of spectral sensitivity (S=0.1*Smax): 215-325

Visible blindness (Smax/S400nm: 100000

Ordering Information

ERYCA

 UVI Sensor: 215 - 325 nm

Data Sheet

PDFERYCA data sheet - pdf (93kB)