Please contact us with any
questions not covered below. 1. What do UVA/UVB/UVC mean?
2. What is the response time
for the TW30DY?
3. Is it possible to make a
detector with an active area of 32 sq mm (4mm x 8mm)?
4. Do I need anything more to
make a measurement with a UV Sensor?
5. Do I need anything more to
make a measurement with a UV Probe?
6. How can I find out the
relationship between input UV power and output sensor current for an individual
UV sensor?
7. How can I find out the
relationship between input UV power and output sensor current for an individual
UV probe?
8. How fast are the sensors?
|
| 1. What do
UVA/UVB/UVC mean?
The UV region covers the wavelength range 100-400 nm and is
divided into three bands:
UVA
(315-400 nm)
UVB
(280-315 nm)
UVC
(100-280 nm)
|
|
2. I am
interested in the TW30DY UV photodiode. What is the response time for the
device?
The response time for the TiO2 photodiodes is a function of the
radiation intensity. This is due to the polycrystalline nature of the
semiconductor: The TW30DY (as well as TW30DY2 and TW30SY) behaves as given by
the following table:
|
UV Intensity |
Response time |
|
50 W/cm² |
0.06 ms |
|
10 mW/cm² |
0.7 ms |
|
0.5 mW/cm² |
40 ms |
|
0.01 mW/cm² |
500 ms |
If you need faster UV-photodiodes, we recommend the SiC
UV-photodiodes, however they have a significantly smaller photoactive
area. |
|
3. I am
interested in the TW30DY UV photodiode. Is it possible to make a detector with
an active area of 32 sq mm (4mm x 8mm)?
A large area is possible to meet your requirements but would
carry a development cost. |
|
4. Do I need
anything more to make a measurement with a UV Sensor?
UV sensors such as the SG01S, TW30SX and the AG38S-TO, are all
photodiodes. They convert the incoming UV light into a very small current (of
the order of nA). It is not possible to measure this current using typical
bench multimeters so a transimpedence amplifier is required to convert this
current into a usable voltage. The cost of a suitable amplifier and associated
parts can be less than £1.
(It is possible to use the UV sensors in voltage mode and make
measurements using multimeter set to the mV range. However, this is not
recommended as the output is very non linear when in voltage mode.)
Note: The sensor itself does not require a power supply as it
acts as a solar cell generating electricity from the incoming UV radiation. The
transimpedance amplifier will require a power source though. |
|
5. Do I need
anything more to make a measurement with a UV Probe?
The UV Probes either include just a sensor or combine a sensor
with an amplifier. The sensor only parts require a transimpedance amplifier as
per the above question. The UV probes with a sensor and amplifier only require
a power supply and a method of monitoring the output (such as a multimeter).
The part UV_Air_ABC_AMP4-20mA_cable for example, only requires a 24V power
supply to drive the probe and an ammeter to monitor the 4mA to 20mA output
current. |
|
6. How can I find
out the relationship between input UV power and output sensor current for an
individual UV sensor?
The relationship between the input UV power and output current
is given on the data sheet for the wavelength of peak sensitivity. To find out
the value at other wavelengths, the graph of spectral response can be used
(also on the data sheet). These figures are for typical parts as there will be
variation in the sensitivities from part to part. The SiC parts (SG01S etc)
have a variation from part to part of +/-10% compared to the data sheet figure.
For the TiO2 parts (TW30SX etc) the figures given are minimum values with the
actual values being upto 50% greater.
If greater accuracy is required, the sensor can be compared
with a calibrated sensor to get an absolute reading or it can be supplied
calibrated at an extra cost.
Once calibrated at one power level, measurements at other power
levels can be made without further calibration steps as the output is very
linear over many orders of magnitude. |
|
7. How can I find
out the relationship between input UV power and output sensor current for an
individual UV probe?
For UV probes which consist only of a sensor, the above answer
applies. For UV probes with both a sensor and an amplifier built into them,
there is the additional question of the gain of the amplifier. This is normally
factory set to meet the customers requirements so will differ from application
to application. Again the entire probe can be supplied calibrated for an
additional cost. |
|
8. How fast are
the sensors?
Rise and fall times for the most popular sensors are as
follows:
|
SGO1S |
8ns |
|
SGO1M |
25ns |
|
SGO1L |
65ns |
|
AlGaN-UVB |
10ns |
|
TW30SX |
10 µs @ 50 W/cm² 700 µs @ 10
mW/cm² 20 ms @ 1 µW/cm² 200 ms @ 50 nW/cm² 500 ms
@ 1 nW/cm²
Note: fall time given, rise times are faster |
|
| |